AA-7000 Series - Shimadzu

C122-E058A
AA-7000 Series
Shimadzu
Atomic Absorption Spectrophotometers
JQA-0376
Founded in 1875, Shimadzu Corporation, a leader in the
development of advanced technologies, has a distinguished
history of innovation built on the foundation of contributing to
society through science and technology. We maintain a global
network of sales, service, technical support and applications
centers on six continents, and have established long-term
relationships with a host of highly trained distributors located
in over 100 countries. For information about Shimadzu, and
to contact your local office, please visit our Web site at
www.shimadzu.com
SHIMADZU CORPORATION. International Marketing Division
3. Kanda-Nishikicho 1-chome, Chiyoda-ku, Tokyo 101-8448, Japan
Phone: 81(3)3219-5641 Fax. 81(3)3219-5710
URL http://www.shimadzu.com
Printed in Japan 3295-01910-30ANS
The contents of this brochure are subject to change without notice.
AA - 7000 Series
Shimadzu Atomic Absorption Spectrophotometers
Reaching Even Greater Heights
Enhanced Flame Analysis
World-Class High-Sensitivity Furnace
Improved Dual Atomizer System
Advanced Safety Technology
Contents
P 06 - Enhanced Flame Analysis
P 18 - Supports System Management and
Accuracy Management
P 08 - World-Class High-Sensitivity Furnace
P 20 - Standard Parts / PC / Software
P 10 - Improved Dual Atomizer System
P 21 - Optional Accessories / Related Products
P 12 - Advanced Safety Technology
P 24 - Specifications
P 14 -Dual-Background Correction Functions
P 26 - Installation Conditions
P 16 - Clear, Easy-to-Use Software
AA - 7000 Series
Shimadzu Atomic Absorption Spectrophotometers
Reaching Even Greater Heights
Enhanced Flame Analysis
World-Class High-Sensitivity Furnace
Improved Dual Atomizer System
Advanced Safety Technology
Contents
P 06 - Enhanced Flame Analysis
P 18 - Supports System Management and
Accuracy Management
P 08 - World-Class High-Sensitivity Furnace
P 20 - Standard Parts / PC / Software
P 10 - Improved Dual Atomizer System
P 21 - Optional Accessories / Related Products
P 12 - Advanced Safety Technology
P 24 - Specifications
P 14 -Dual-Background Correction Functions
P 26 - Installation Conditions
P 16 - Clear, Easy-to-Use Software
System Configuration can Evolve with Your Needs
The AA-7000 Series can be upgraded by adding units to allow the system to handle the analysis targets.
ASC
*1
AA-7000F Manual Burner System
AUTO
AAC
AA-7000F
AAC
GFA
ASC
*2
*3
AA-7000F Auto Burner System
AA-7000F Dual Atomizer System
AAC
ASC
GFA
*2
AA-7000G
GFA-7000
Major Fields of Application
Graphite Furnace
Atomizer
ASC
Autosampler
AUTO
Dual Atomizer System
Automatic adjustment of atomizer
Automatic serach of burner height
*1 Requires ASC stand kit.
*2 Requires ASK-7000.
*3 Automatic burner height search function can be used.
AA-7000G Auto Sampler System
Flame Selection
Analysis Sensitivities
Flame and Measurement Procedures
Furnace
Flame
AA-7000 Series supports a wide range of
analysis applications.
Auto Atomizer
Changer
Air-C2H2 N2O-C2H2 HVG MVU
Concentration limit of detection
Environment
Seawater, river water, effluent,
sludge, air-borne dust
Metals, Semiconductors, Ceramics
Metals, minerals, glass,
ceramics, IC chips
Petroleum, Chemicals, Polymers
Petroleum, oil, catalysts,
chemical products, biodiesel
Medical, Biology, Pharmaceuticals
Blood, animals, plants,
drugs, food products
(ppm)
H
He
H
He
Li Be
B
C
N
O
F Ne
Li Be
B
C
N
O
F Ne
Na Mg
Al Si
P
S
Cl Ar
Na Mg
Al Si
P
S
Cl Ar
K Ca Sc Ti
Rb Sr
Y
V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr
Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te
I
Xe
K Ca Sc Ti
Rb Sr
Y
AA-7000F Flame Model
V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr
Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te
I
Xe
Cs Ba La Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn
Cs Ba La Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn
Fr Ra Ac
Fr Ra Ac
Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
Th Pa U Np Pu Am Cm Bk Cf Es Fm Md No Lr
Th Pa U Np Pu Am Cm Bk Cf Es Fm Md No Lr
* May differ according to coexisting substances in the sample.
4
(ppb)
AA-7000G Furnace Model (with options attached)
5
System Configuration can Evolve with Your Needs
The AA-7000 Series can be upgraded by adding units to allow the system to handle the analysis targets.
ASC
*1
AA-7000F Manual Burner System
AUTO
AAC
AA-7000F
AAC
GFA
ASC
*2
*3
AA-7000F Auto Burner System
AA-7000F Dual Atomizer System
AAC
ASC
GFA
*2
AA-7000G
GFA-7000
Major Fields of Application
Graphite Furnace
Atomizer
ASC
Autosampler
AUTO
Dual Atomizer System
Automatic adjustment of atomizer
Automatic serach of burner height
*1 Requires ASC stand kit.
*2 Requires ASK-7000.
*3 Automatic burner height search function can be used.
AA-7000G Auto Sampler System
Flame Selection
Analysis Sensitivities
Flame and Measurement Procedures
Furnace
Flame
AA-7000 Series supports a wide range of
analysis applications.
Auto Atomizer
Changer
Air-C2H2 N2O-C2H2 HVG MVU
Concentration limit of detection
Environment
Seawater, river water, effluent,
sludge, air-borne dust
Metals, Semiconductors, Ceramics
Metals, minerals, glass,
ceramics, IC chips
Petroleum, Chemicals, Polymers
Petroleum, oil, catalysts,
chemical products, biodiesel
Medical, Biology, Pharmaceuticals
Blood, animals, plants,
drugs, food products
(ppm)
H
He
H
He
Li Be
B
C
N
O
F Ne
Li Be
B
C
N
O
F Ne
Na Mg
Al Si
P
S
Cl Ar
Na Mg
Al Si
P
S
Cl Ar
K Ca Sc Ti
Rb Sr
Y
V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr
Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te
I
Xe
K Ca Sc Ti
Rb Sr
Y
AA-7000F Flame Model
V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr
Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te
I
Xe
Cs Ba La Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn
Cs Ba La Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn
Fr Ra Ac
Fr Ra Ac
Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
Th Pa U Np Pu Am Cm Bk Cf Es Fm Md No Lr
Th Pa U Np Pu Am Cm Bk Cf Es Fm Md No Lr
* May differ according to coexisting substances in the sample.
4
(ppb)
AA-7000G Furnace Model (with options attached)
5
Enhanced Flame Analysis
Advantages of the Double-Beam System
Long-term stability
Monochromator
Detector
High sensitivity
The graph shows the direct
measurement results for 0.1В ppm lead (Pb).
Absorbance
The graph shows the results of
measurements on 2В ppm copper (Cu)
conducted over at least one hour. (The
plot shows mean values for 11 repeated
measurements.) Over the course of
more than 600 measurements, the
instrument achieved a relative standard
deviation within 1%.
A
b
s
Measurement time
Conc(ppm)
Data Showing Long-Term Stability for
Flame Cu Measurements
0.1В ppm to 0.4В ppm Calibration Curve for Lead (Pb)
Support for Developing Analytical Conditions
Automatic gas flow rate optimization
Chopper mirror
Automatic searching for optimal fuel gas flow rate (Japanese Patent 2099886). It is important to determine
the optimal gas flow rate for the flame when using an organic solvent or after changing the burner height.
The AA-7000F Series automatically optimizes the gas flow rate by measuring the changes in absorbance
between a blank and a standard sample. The difference between the two is displayed on the screen. The
gas flow rate achieving the highest sensitivity is detected and this value is automatically set as the gas
flow rate value.
Reference beam
D2 lamp
Sample beam
Beam splitter
Automatic burner height optimization (with options attached)
The absorption sensitivity for flame analysis is also affected by the burner height. This results from
variations in the flame temperature due to the burner height. The flame temperature is also affected by the
matrix components. The optional AAC-7000 adjusts the burner height in 0.5В mm steps to automatically
search for the optimal conditions.
Optional Autosampler Reduces Analysis Workload
Burner head (or graphite tube)
Low carryover
Great care must be taken to avoid carryover during flame analysis. The new ASC-7000
autosampler rinses the nozzle at the rinsing port on the overflow mechanism after each
sample measurement is complete. This ensures 10-4 max. carryover during the
measurements of multiple samples. The graph shows the results of consecutive analyses
of 10В ppm, 20В ppm, and 50В ppm sodium (Na) standard solutions in the EMISSION mode.
No carryover could be detected when measuring the 10В ppm standard sample
immediately after the measuring the 50В ppm Na standard sample.
Hollow cathode lamp
Discharge
Conventional method
Overflow method
Data showing low carryover
Trace Sample Analysis Using Micro Sampling
Micro sampling
At least 1В mL (1000В ОјL) volume of a liquid sample is required for the continuous intake of sample during normal flame analysis. With micro sampling,
however, approximately 50 to 90В ОјL of sample is injected in one shot into the flame and quantitation is based on the height and area of the peak signal
obtained. This method offers the advantages below.
Advantages
Double-Beam Optics and Stable Hardware Achieve Superior Stability
The AA-7000 Series incorporates newly developed 3D double-beam optics.
The optical system has been designed to produce its maximum performance for each measurement method through optimal
adjustment of the light beam and light beam digital filter, and by using optical components that restrict light losses.
6
. Permits analysis of small sample volumes
. Multi-element analysis of small sample volumes
. No blockage of burner slot with samples having a high salt concentration
. Synchronized with the autosampler for auto-dilution measurements
* Synchronization with the autosampler requires the optional micro sampling kit.
Example of micro sampling analysis
Sampling port
Solenoid
valve
(closed)
1. Sample injected into
micro sampling port.
Sampling port
Solenoid
valve
(opened)
2. Solenoid valve opens to inject sample
in one shot into the flame.
Examples of the measured waveforms (overlaid) and calibration curve for micro sampling
analysis are shown to the right.
A 2-ppm Cu standard solution was prepared by auto-dilution using the autosampler.
The autosampler can also be used to conduct dilution measurements of the sample.
(Injection volume: 90В ОјL)
7
Enhanced Flame Analysis
Advantages of the Double-Beam System
Long-term stability
Monochromator
Detector
High sensitivity
The graph shows the direct
measurement results for 0.1В ppm lead (Pb).
Absorbance
The graph shows the results of
measurements on 2В ppm copper (Cu)
conducted over at least one hour. (The
plot shows mean values for 11 repeated
measurements.) Over the course of
more than 600 measurements, the
instrument achieved a relative standard
deviation within 1%.
A
b
s
Measurement time
Conc(ppm)
Data Showing Long-Term Stability for
Flame Cu Measurements
0.1В ppm to 0.4В ppm Calibration Curve for Lead (Pb)
Support for Developing Analytical Conditions
Automatic gas flow rate optimization
Chopper mirror
Automatic searching for optimal fuel gas flow rate (Japanese Patent 2099886). It is important to determine
the optimal gas flow rate for the flame when using an organic solvent or after changing the burner height.
The AA-7000F Series automatically optimizes the gas flow rate by measuring the changes in absorbance
between a blank and a standard sample. The difference between the two is displayed on the screen. The
gas flow rate achieving the highest sensitivity is detected and this value is automatically set as the gas
flow rate value.
Reference beam
D2 lamp
Sample beam
Beam splitter
Automatic burner height optimization (with options attached)
The absorption sensitivity for flame analysis is also affected by the burner height. This results from
variations in the flame temperature due to the burner height. The flame temperature is also affected by the
matrix components. The optional AAC-7000 adjusts the burner height in 0.5В mm steps to automatically
search for the optimal conditions.
Optional Autosampler Reduces Analysis Workload
Burner head (or graphite tube)
Low carryover
Great care must be taken to avoid carryover during flame analysis. The new ASC-7000
autosampler rinses the nozzle at the rinsing port on the overflow mechanism after each
sample measurement is complete. This ensures 10-4 max. carryover during the
measurements of multiple samples. The graph shows the results of consecutive analyses
of 10В ppm, 20В ppm, and 50В ppm sodium (Na) standard solutions in the EMISSION mode.
No carryover could be detected when measuring the 10В ppm standard sample
immediately after the measuring the 50В ppm Na standard sample.
Hollow cathode lamp
Discharge
Conventional method
Overflow method
Data showing low carryover
Trace Sample Analysis Using Micro Sampling
Micro sampling
At least 1В mL (1000В ОјL) volume of a liquid sample is required for the continuous intake of sample during normal flame analysis. With micro sampling,
however, approximately 50 to 90В ОјL of sample is injected in one shot into the flame and quantitation is based on the height and area of the peak signal
obtained. This method offers the advantages below.
Advantages
Double-Beam Optics and Stable Hardware Achieve Superior Stability
The AA-7000 Series incorporates newly developed 3D double-beam optics.
The optical system has been designed to produce its maximum performance for each measurement method through optimal
adjustment of the light beam and light beam digital filter, and by using optical components that restrict light losses.
6
. Permits analysis of small sample volumes
. Multi-element analysis of small sample volumes
. No blockage of burner slot with samples having a high salt concentration
. Synchronized with the autosampler for auto-dilution measurements
* Synchronization with the autosampler requires the optional micro sampling kit.
Example of micro sampling analysis
Sampling port
Solenoid
valve
(closed)
1. Sample injected into
micro sampling port.
Sampling port
Solenoid
valve
(opened)
2. Solenoid valve opens to inject sample
in one shot into the flame.
Examples of the measured waveforms (overlaid) and calibration curve for micro sampling
analysis are shown to the right.
A 2-ppm Cu standard solution was prepared by auto-dilution using the autosampler.
The autosampler can also be used to conduct dilution measurements of the sample.
(Injection volume: 90В ОјL)
7
World-Class High-Sensitivity Furnace
High-Performance Autosampler
Mix and inject up to four sample types (diluted solution, standard solution,
sample, matrix modifier, etc.). (Of course, unmixed samples can also be injected.)
Select a fluororesin tube or pipette tip as the injection nozzle.
Automatic dilution and re-analysis if the calibration curve concentration is exceeded.
Example of Automatic Dilution and Re-Analysis
Analysis of cadmium (Cd): The maximum concentration of the calibration curve has been set to 1ppb. At 1.8ppb, the sample of [003] has exceeded the
maximum concentration. As a resultit has been automatically diluted by 4 times and re-measured.
High-Sensitivity Application Data (Analysis of Selenium (Se))
Independent control of the gas flow through the graphite tube during
atomization achieves extremely high-sensitivity measurements.
(Japanese Patent 2067563)
Measurement Example Showing High Sensitivity
Analysis of 0 to 5В ppb selenium (Se): Sensitivity is adequate for 1В ppb measurements.
(20В ОјL injected volume, Pb modifier, pyro-coated graphite tube)
Digital temperature control and electronic gas flow control enhance stability
Combination of a high-sensitivity optical sensor and unique digital temperature control
technology achieves highly accurate temperature control in all temperature regions from
drying to atomization. An electronic flow controller can precisely control the inner gas flow rate
in 0.01В L/min units. These control techniques significantly enhance both the sensitivity and the
data stability.
Measurement Example Showing High Stability
Analysis of 0.5В ppb lead (Pb): Approx. 5% relative standard deviation
after 5 repeated measurements is adequate for quantitative analysis.
Selecting the Graphite Tube
High-density graphite tube
Achieves World-Class Lower Limits of Detection
Advances in optics and a new graphite furnace design achieve improved lower limits of
detection for furnace analysis (comparison with previous Shimadzu models).
This superb analysis performance is possible in all fields.
Enhanced Sensitivity due to Graphite Furnace
Previous
Pb 0.08ppb
Mn 0.02ppb
8
New
0 .0 5 p p b
(P/N 206-50587)
Can be used for all elements. Especially
effective for low-boiling point elements
(Cd, Pb, Na, K, Zn, Mg, etc.).
Pyro-coated graphite tube
(P/N 206-50588)
Effective for elements that readily form
carbides (Ni, Fe, Cu, Ca, Ti, Si, V, Mo, etc.).
Platform tube
(P/N 206-50887-02)
Restricts chemical interference due to
coexisting substances. Effective for the analysis
of environmental samples and biosamples,
such as sea water and industrial waste.
* Depending on the state of the sample, some other combinations may be appropriate.
0 .0 1 p p b
9
World-Class High-Sensitivity Furnace
High-Performance Autosampler
Mix and inject up to four sample types (diluted solution, standard solution,
sample, matrix modifier, etc.). (Of course, unmixed samples can also be injected.)
Select a fluororesin tube or pipette tip as the injection nozzle.
Automatic dilution and re-analysis if the calibration curve concentration is exceeded.
Example of Automatic Dilution and Re-Analysis
Analysis of cadmium (Cd): The maximum concentration of the calibration curve has been set to 1ppb. At 1.8ppb, the sample of [003] has exceeded the
maximum concentration. As a resultit has been automatically diluted by 4 times and re-measured.
High-Sensitivity Application Data (Analysis of Selenium (Se))
Independent control of the gas flow through the graphite tube during
atomization achieves extremely high-sensitivity measurements.
(Japanese Patent 2067563)
Measurement Example Showing High Sensitivity
Analysis of 0 to 5В ppb selenium (Se): Sensitivity is adequate for 1В ppb measurements.
(20В ОјL injected volume, Pb modifier, pyro-coated graphite tube)
Digital temperature control and electronic gas flow control enhance stability
Combination of a high-sensitivity optical sensor and unique digital temperature control
technology achieves highly accurate temperature control in all temperature regions from
drying to atomization. An electronic flow controller can precisely control the inner gas flow rate
in 0.01В L/min units. These control techniques significantly enhance both the sensitivity and the
data stability.
Measurement Example Showing High Stability
Analysis of 0.5В ppb lead (Pb): Approx. 5% relative standard deviation
after 5 repeated measurements is adequate for quantitative analysis.
Selecting the Graphite Tube
High-density graphite tube
Achieves World-Class Lower Limits of Detection
Advances in optics and a new graphite furnace design achieve improved lower limits of
detection for furnace analysis (comparison with previous Shimadzu models).
This superb analysis performance is possible in all fields.
Enhanced Sensitivity due to Graphite Furnace
Previous
Pb 0.08ppb
Mn 0.02ppb
8
New
0 .0 5 p p b
(P/N 206-50587)
Can be used for all elements. Especially
effective for low-boiling point elements
(Cd, Pb, Na, K, Zn, Mg, etc.).
Pyro-coated graphite tube
(P/N 206-50588)
Effective for elements that readily form
carbides (Ni, Fe, Cu, Ca, Ti, Si, V, Mo, etc.).
Platform tube
(P/N 206-50887-02)
Restricts chemical interference due to
coexisting substances. Effective for the analysis
of environmental samples and biosamples,
such as sea water and industrial waste.
* Depending on the state of the sample, some other combinations may be appropriate.
0 .0 1 p p b
9
Improved Dual Atomizer System
New Atomizer Design Permits Easy Switching Between Flame and Furnace Measurements
The new drive mechanism halves the switching time between flame and furnace measurements
(comparison with previous Shimadzu models).
The atomizer unit is switched automatically by software
operation. No troublesome replacement of the atomizer unit
is required. No need to disconnect pipes or wires.
Flame measurement: Burner head
intersects the light axis (red line).
Furnace measurement: Furnace
intersects the light axis (red line).
A single autosampler can be used for both flame and
furnace measurements. Simply slide it left or right. No need
to purchase two autosamplers.
Flame measurement
Furnace measurement
Slim, Easy-to-Use Design
Smallest footprint for a full system
The system with a dual atomizer fits into a 960 mm width
(including power supply for the graphite furnace atomizer),
achieving the world's smallest installation footprint* for a full
system.
Easily attach an autosampler
Direct connection of the autosampler to the
AA instrument minimizes the installation
footprint and simplifies maintenance.
* Shimadzu in-house investigation. Current as of November 2008.
Lamp house
Just slide the cover to the side for easy
replacement of the hollow cathode lamps
through the large opening.
AA-6300
Single Unit Handles High Concentration to Trace Analysis Samples
When attached to the AA-7000F, the optional AAC-7000 Auto Atomizer Changer* creates an integral burner and furnace that
remains permanently installed in the combustion chamber. It supports both flame analysis of high concentration samples and
furnace analysis of trace samples, without changing the atomizer unit. The integral construction with the atomizer saves space.
* The AAC-7000 cannot be attached to an AA-7000G instrument.
10
AA-7000F Dual Atomizer System
* Dimensions do not include ASC-7000. Height (H) does not include protrusions.
11
Improved Dual Atomizer System
New Atomizer Design Permits Easy Switching Between Flame and Furnace Measurements
The new drive mechanism halves the switching time between flame and furnace measurements
(comparison with previous Shimadzu models).
The atomizer unit is switched automatically by software
operation. No troublesome replacement of the atomizer unit
is required. No need to disconnect pipes or wires.
Flame measurement: Burner head
intersects the light axis (red line).
Furnace measurement: Furnace
intersects the light axis (red line).
A single autosampler can be used for both flame and
furnace measurements. Simply slide it left or right. No need
to purchase two autosamplers.
Flame measurement
Furnace measurement
Slim, Easy-to-Use Design
Smallest footprint for a full system
The system with a dual atomizer fits into a 960 mm width
(including power supply for the graphite furnace atomizer),
achieving the world's smallest installation footprint* for a full
system.
Easily attach an autosampler
Direct connection of the autosampler to the
AA instrument minimizes the installation
footprint and simplifies maintenance.
* Shimadzu in-house investigation. Current as of November 2008.
Lamp house
Just slide the cover to the side for easy
replacement of the hollow cathode lamps
through the large opening.
AA-6300
Single Unit Handles High Concentration to Trace Analysis Samples
When attached to the AA-7000F, the optional AAC-7000 Auto Atomizer Changer* creates an integral burner and furnace that
remains permanently installed in the combustion chamber. It supports both flame analysis of high concentration samples and
furnace analysis of trace samples, without changing the atomizer unit. The integral construction with the atomizer saves space.
* The AAC-7000 cannot be attached to an AA-7000G instrument.
10
AA-7000F Dual Atomizer System
* Dimensions do not include ASC-7000. Height (H) does not include protrusions.
11
Advanced Safety Technology
Safety Functions for Flame-Specification Instruments
World first! –– flame extinguished automatically by vibration sensor
The flame is automatically extinguished when the new built-in sensor detects vibrations.
Instruments are completely safe during severe shaking caused by earthquakes.
Multimode automatic gas leak check
The fuel gas pipes inside the instrument gas controller are automatically checked
for gas leaks when the power is turned on and when the flame is extinguished.
Should a gas leak be detected, a warning buzzer sounds and a warning is
displayed on the screen.
Automatic flame ignition and extinguishing
Flame ignition and extinguishing operations are extremely simple. The Air-C2H2
flame priority ignition mechanism prevents flashback.
Automatic Air-N2O switching system with acetylene flow-rate monitor
After the ignition of an Air-C2H2 flame, the flame automatically switches to an
N2O-C2H2 flame.
If the C2H2 flow rate does not increase due to solenoid valve trouble, for example,
switching to the support gas is disabled to prevent flashback.
Pressure monitor prevents flashback
The connected options and operation of safety mechanisms are
checked during instrument initialization.
Automatic gas shut-off by flame monitor
Sudden power interruption detection and re-ignition safety measures
Drain tank level monitor
Mechanism to prevent misuse of the burner
Safety Functions for Furnace-Specification Instruments
Cooling water flow rate monitor
Argon gas pressure monitor
Furnace cooling check
Overcurrent protection unit (double-checked by circuit protector and optical sensor)
Safety-Conscious Design
Flame-retardant materials
UL 94V-0-rated flame-retardant materials are
used for the instrument external covers and
atomizer unit. This superior design pays due
consideration to safety.
Durable pipes and reliable couplings
Comprehensive Safety Functions in Every System
Safety is an extremely important consideration for flame atomic absorption spectrophotometers using acetylene gas. The AA-7000
Series instruments are the first* ones in the world fitted with a vibration sensor as standard. They also incorporate a comprehensive
range of safety mechanisms, including gas leak detectors.
All pipes are selected for durability and
used with highly reliable couplings.
Safe ignition switches
The switches perfectly fit finger contours and are
positioned near the window.
To prevent inadvertent operation, both switches
must be pressed simultaneously to start ignition.
Clear window
The large window permits observation of the
flame, and the burner can be easily accessed
through the large opening.
The window closes when released to eliminate
the chance of it being left open.
* Shimadzu in-house investigation. Current as of November 2008.
12
13
Advanced Safety Technology
Safety Functions for Flame-Specification Instruments
World first! –– flame extinguished automatically by vibration sensor
The flame is automatically extinguished when the new built-in sensor detects vibrations.
Instruments are completely safe during severe shaking caused by earthquakes.
Multimode automatic gas leak check
The fuel gas pipes inside the instrument gas controller are automatically checked
for gas leaks when the power is turned on and when the flame is extinguished.
Should a gas leak be detected, a warning buzzer sounds and a warning is
displayed on the screen.
Automatic flame ignition and extinguishing
Flame ignition and extinguishing operations are extremely simple. The Air-C2H2
flame priority ignition mechanism prevents flashback.
Automatic Air-N2O switching system with acetylene flow-rate monitor
After the ignition of an Air-C2H2 flame, the flame automatically switches to an
N2O-C2H2 flame.
If the C2H2 flow rate does not increase due to solenoid valve trouble, for example,
switching to the support gas is disabled to prevent flashback.
Pressure monitor prevents flashback
The connected options and operation of safety mechanisms are
checked during instrument initialization.
Automatic gas shut-off by flame monitor
Sudden power interruption detection and re-ignition safety measures
Drain tank level monitor
Mechanism to prevent misuse of the burner
Safety Functions for Furnace-Specification Instruments
Cooling water flow rate monitor
Argon gas pressure monitor
Furnace cooling check
Overcurrent protection unit (double-checked by circuit protector and optical sensor)
Safety-Conscious Design
Flame-retardant materials
UL 94V-0-rated flame-retardant materials are
used for the instrument external covers and
atomizer unit. This superior design pays due
consideration to safety.
Durable pipes and reliable couplings
Comprehensive Safety Functions in Every System
Safety is an extremely important consideration for flame atomic absorption spectrophotometers using acetylene gas. The AA-7000
Series instruments are the first* ones in the world fitted with a vibration sensor as standard. They also incorporate a comprehensive
range of safety mechanisms, including gas leak detectors.
All pipes are selected for durability and
used with highly reliable couplings.
Safe ignition switches
The switches perfectly fit finger contours and are
positioned near the window.
To prevent inadvertent operation, both switches
must be pressed simultaneously to start ignition.
Clear window
The large window permits observation of the
flame, and the burner can be easily accessed
through the large opening.
The window closes when released to eliminate
the chance of it being left open.
* Shimadzu in-house investigation. Current as of November 2008.
12
13
Dual-Background Correction Functions
The optimal background correction methods are installed as standard: high-speed self-reversal
These functions can correct for spectral interference in flame measurement.
Selecting the optimal background correction method for each sample ensures
accurate and reliable analysis results.
Samples suitable for the SR method
Samples suitable for the D 2 method
Samples with a complex matrix
Purified water, tap water, environmental water, etc.
(Containing a large quantity of a specific element as the main component)
Samples with a relatively simple matrix
SR (high-speed self-reversal) method –– accurate background correction over a wide range
Features
1. High-speed self-reversal (SR) correction is generally more accurate than deuterium lamp (D2) correction. As both atomic absorption and background absorption can be
measured using a single lamp, the correction errors due to light-axis misalignment are extremely small. This is ideal for the quantitation of trace components in a matrix
exhibiting complex background absorption, such as bio-samples and metals.
2. Permits background correction over the entire wavelength range from 185 nm to 900 nm.
3. This method can correct for spectral interference due to neighboring lines that can occur when a resonance line for another element exists near the analytical line for the
target element. (See table on next page.)
4. As no polarizer is used, measurements are possible with low light losses and a high S/N ratio.
method (SR method) and deuterium lamp method (D2 method).
Examples of elements and wavelengths causing spectral
interference problems due to neighboring lines
Measured Element Analytical Line (nm) Coexisting Element Absorption Line (nm)
309.28
Mg
309.30
193.76
Fe
193.73
422.67
Fe
422.64
Cd
228.80
Ni
228.84
Cu
324.75
Fe
324.73
Ga
294.36
Ni
294.39
Mg
285.21
Fe
285.18
Ni
232.00
Fe
232.04
Pb
217.00
Fe
216.95
Sb
217.58
Fe
217.55
Se
196.03
Fe
196.05
Si
251.61
Fe
251.69
Zn
213.856
Fe
213.8589
Al
As
Ca
The SR method is suitable for samples containing elements that cause
problems with spectral interference, as shown in the table to the left.
D2 (deuterium lamp) method –– highly sensitive background correction
Features
1. Detection sensitivity is superior to the SR method. Therefore, this method is suitable for the analysis of samples with a simple matrix requiring high sensitivity, such as the
measurement of trace levels of impurities in ultrapure water or environmental analyses.
2. As the lighting frequency is higher than with the SR method, it can eliminate noise due to emission components of the flame or graphite tube to permit accurate atomic
absorption measurements.
3. The original hollow cathode lamp can be used.
D2 lamp spectrum
Principle
Background absorption
5. The rapid lamp lighting permits accurate measurement unaffected by emission noise in the atomizer.
* Hollow cathode lamp L-2433 is required to use the SR method. Hollow cathode lamp L-2433 can also be used for the D2 method.
IH spectrum
Lamp current
Background
Atomic absorption
Atomic
absorption
IH
Wavelength
IL
Wavelength
Absorption
layer
IL spectrum
Atomic
absorption
Background
Time
Atomic absorption
(Lamp energy)
(Sample measurement)
(Energy component ratio)
Atomic absorption
Principle
Background absorption
A small current IL (approx. 10 mA) and a large current IH (approx. 500 mA)
are alternately passed through the hollow cathode lamp. The lamp
emission spectrum when the large current flows has a depression in the
center (self-reverse), due to self-absorption of the large number of
sputtered atoms in the atom cloud, as shown in the diagram to the left. No
significant atomic absorption is apparent and background absorption
mainly occurs. Conversely, the lamp emission spectrum when the small
current flows comprises a single narrow peak resulting from both atomic
absorption and background absorption. By determining the difference
between the two types of absorption, it is possible to accurately correct for
the background absorption and measure the true atomic absorption.
Hollow cathode lamp spectrum
Atomic absorption signal
Background signal
Atomic absorption + background absorption
Examples suitable for D2 method (where no difference results between SR and D2 methods)
Example: Measurement of trace levels of lead in 2% NaCl solution by molecular absorption (analysis of Pb in 2% NaCl solution)
BGC-D 2 method
BGC-SR method
Examples suitable for SR method (where differences result between SR and D2 methods)
Example: Measurement of trace levels of zinc in iron (analysis of Zn in Fe solution)
BGC-SR method
The deuterium lamp method involves lighting the hollow cathode lamp and the deuterium
lamp alternately at high speed. After separation by the monochromator, the light from the
deuterium lamp has a bandwidth from 0.1 to 5 nm. Therefore, an atomic absorption with a
line width of only about 1/1000 nm is almost unobservable compared to the background
absorption due to wide-bandwidth molecular absorption. However, as the light from the
hollow cathode lamp has approximately the same bandwidth as the atomic absorption
band, the total of the atomic absorption and the background absorption can be observed.
With the deuterium lamp (D2) method, light from both sources passes through the
atomizer. The difference in absorbance is determined to conduct background correction.
BGC-D 2 method
Atomic absorption signal
Background signal
Zn 0.25ppm
Background signal
Zn 0.50ppm
Fe 0.1%
Fe 0.5%
Fe 0.5%
Zn 0.25ppm
Fe 0.5%
Zn 0.5ppm
Fe 0.75%
The identical 0.5 ppm Zn solution is accurately corrected to
the same absorbance at (2) and (6).
14
Due to inadequate correction, the absorbance is higher at
(6) than at (2) for the identical 0.5 ppm Zn solution.
Atomic absorption signal
Zn 0.3ppm
Spike
0ppb
2ppb
4ppb
Spike
0ppb
2ppb
4ppb
It can be seen that the sensitivity is higher with the BGC-D2 method.
15
Dual-Background Correction Functions
The optimal background correction methods are installed as standard: high-speed self-reversal
These functions can correct for spectral interference in flame measurement.
Selecting the optimal background correction method for each sample ensures
accurate and reliable analysis results.
Samples suitable for the SR method
Samples suitable for the D 2 method
Samples with a complex matrix
Purified water, tap water, environmental water, etc.
(Containing a large quantity of a specific element as the main component)
Samples with a relatively simple matrix
SR (high-speed self-reversal) method –– accurate background correction over a wide range
Features
1. High-speed self-reversal (SR) correction is generally more accurate than deuterium lamp (D2) correction. As both atomic absorption and background absorption can be
measured using a single lamp, the correction errors due to light-axis misalignment are extremely small. This is ideal for the quantitation of trace components in a matrix
exhibiting complex background absorption, such as bio-samples and metals.
2. Permits background correction over the entire wavelength range from 185 nm to 900 nm.
3. This method can correct for spectral interference due to neighboring lines that can occur when a resonance line for another element exists near the analytical line for the
target element. (See table on next page.)
4. As no polarizer is used, measurements are possible with low light losses and a high S/N ratio.
method (SR method) and deuterium lamp method (D2 method).
Examples of elements and wavelengths causing spectral
interference problems due to neighboring lines
Measured Element Analytical Line (nm) Coexisting Element Absorption Line (nm)
309.28
Mg
309.30
193.76
Fe
193.73
422.67
Fe
422.64
Cd
228.80
Ni
228.84
Cu
324.75
Fe
324.73
Ga
294.36
Ni
294.39
Mg
285.21
Fe
285.18
Ni
232.00
Fe
232.04
Pb
217.00
Fe
216.95
Sb
217.58
Fe
217.55
Se
196.03
Fe
196.05
Si
251.61
Fe
251.69
Zn
213.856
Fe
213.8589
Al
As
Ca
The SR method is suitable for samples containing elements that cause
problems with spectral interference, as shown in the table to the left.
D2 (deuterium lamp) method –– highly sensitive background correction
Features
1. Detection sensitivity is superior to the SR method. Therefore, this method is suitable for the analysis of samples with a simple matrix requiring high sensitivity, such as the
measurement of trace levels of impurities in ultrapure water or environmental analyses.
2. As the lighting frequency is higher than with the SR method, it can eliminate noise due to emission components of the flame or graphite tube to permit accurate atomic
absorption measurements.
3. The original hollow cathode lamp can be used.
D2 lamp spectrum
Principle
Background absorption
5. The rapid lamp lighting permits accurate measurement unaffected by emission noise in the atomizer.
* Hollow cathode lamp L-2433 is required to use the SR method. Hollow cathode lamp L-2433 can also be used for the D2 method.
IH spectrum
Lamp current
Background
Atomic absorption
Atomic
absorption
IH
Wavelength
IL
Wavelength
Absorption
layer
IL spectrum
Atomic
absorption
Background
Time
Atomic absorption
(Lamp energy)
(Sample measurement)
(Energy component ratio)
Atomic absorption
Principle
Background absorption
A small current IL (approx. 10 mA) and a large current IH (approx. 500 mA)
are alternately passed through the hollow cathode lamp. The lamp
emission spectrum when the large current flows has a depression in the
center (self-reverse), due to self-absorption of the large number of
sputtered atoms in the atom cloud, as shown in the diagram to the left. No
significant atomic absorption is apparent and background absorption
mainly occurs. Conversely, the lamp emission spectrum when the small
current flows comprises a single narrow peak resulting from both atomic
absorption and background absorption. By determining the difference
between the two types of absorption, it is possible to accurately correct for
the background absorption and measure the true atomic absorption.
Hollow cathode lamp spectrum
Atomic absorption signal
Background signal
Atomic absorption + background absorption
Examples suitable for D2 method (where no difference results between SR and D2 methods)
Example: Measurement of trace levels of lead in 2% NaCl solution by molecular absorption (analysis of Pb in 2% NaCl solution)
BGC-D 2 method
BGC-SR method
Examples suitable for SR method (where differences result between SR and D2 methods)
Example: Measurement of trace levels of zinc in iron (analysis of Zn in Fe solution)
BGC-SR method
The deuterium lamp method involves lighting the hollow cathode lamp and the deuterium
lamp alternately at high speed. After separation by the monochromator, the light from the
deuterium lamp has a bandwidth from 0.1 to 5 nm. Therefore, an atomic absorption with a
line width of only about 1/1000 nm is almost unobservable compared to the background
absorption due to wide-bandwidth molecular absorption. However, as the light from the
hollow cathode lamp has approximately the same bandwidth as the atomic absorption
band, the total of the atomic absorption and the background absorption can be observed.
With the deuterium lamp (D2) method, light from both sources passes through the
atomizer. The difference in absorbance is determined to conduct background correction.
BGC-D 2 method
Atomic absorption signal
Background signal
Zn 0.25ppm
Background signal
Zn 0.50ppm
Fe 0.1%
Fe 0.5%
Fe 0.5%
Zn 0.25ppm
Fe 0.5%
Zn 0.5ppm
Fe 0.75%
The identical 0.5 ppm Zn solution is accurately corrected to
the same absorbance at (2) and (6).
14
Due to inadequate correction, the absorbance is higher at
(6) than at (2) for the identical 0.5 ppm Zn solution.
Atomic absorption signal
Zn 0.3ppm
Spike
0ppb
2ppb
4ppb
Spike
0ppb
2ppb
4ppb
It can be seen that the sensitivity is higher with the BGC-D2 method.
15
Clear, Easy-to-Use Software
Features of the WizAArd Software
The WizAArd software used with the AA-7000 runs under Windows Vista/XP.
Wizard Functions Make Setting the Conditions So Simple That Even a Novice Can Do It.
Just set the measurement conditions using WizAArd to
complete the general settings.
Software Provides Comprehensive Information Displays and Flexible Settings
Identical software operation for flame and furnace measurements
Even if the calibration curve is created after sample measurements,
concentration calculations can be conducted on past data. This allows
trial measurements to be directly used as actual measurements.
Extrapolation of calibration curve range. If the sample concentration
exceeds the calibration curve range, additional measurements can be
performed on standard solutions to re-calculate the concentration.
The initial settings can also be completed simply
by following the Wizard procedures.
Use existing calibration curves for measurements. Useful for sample
concentration order checks and for checking the instrument sensitivity.
One-touch re-measurement operation. No need to repeat the
measurement from the beginning.
Start Wizard.
Create and save multiple calibration curves on a single worksheet.
Select the appropriate calibration curve for the sample concentration and
composition.
Select element.
Set calibration curve and sample parameters.
Combining a high-performance autosampler simplifies advanced settings
Set monochromator parameters.
Real-time drying and ashing progress display for furnace measurement
Connect to instrument.
A higher temperature cleaning operation than the cleaning stage can be
added for each measurement. This reduces the graphite tube memory
effect. Tube deterioration is reduced by minimizing the cleaning
operations.
Measurement Screen Layout Shows Measurement Status at a Glance
Measured element
Real-time signal monitor
Signal profile display
Calibration curve display
Add optional autosampler rinsing operations to ensure adequate rinsing
of the autosampler after completing the measurement of
high-concentration samples.
Set up sample dilution and spike-and-recovery tests on the same screen.
MRT (Measured Results Table)
The worksheet shows sample names,
absorbance, concentrations,
and correction calibration results.
Overlay function allows size and shape comparison of peak profiles.
Large-volume injection permits high-sensitivity measurements.
Precision control of drying and ashing temperatures reduces analysis time.
Lamp History Function Assists with Lamp Management
The accumulated operation time for each lamp displayed on the lamp
registration screen assists with lamp service life management. Multiple lamps
for the same element are differentiated using lamp IDs.
16
17
Clear, Easy-to-Use Software
Features of the WizAArd Software
The WizAArd software used with the AA-7000 runs under Windows Vista/XP.
Wizard Functions Make Setting the Conditions So Simple That Even a Novice Can Do It.
Just set the measurement conditions using WizAArd to
complete the general settings.
Software Provides Comprehensive Information Displays and Flexible Settings
Identical software operation for flame and furnace measurements
Even if the calibration curve is created after sample measurements,
concentration calculations can be conducted on past data. This allows
trial measurements to be directly used as actual measurements.
Extrapolation of calibration curve range. If the sample concentration
exceeds the calibration curve range, additional measurements can be
performed on standard solutions to re-calculate the concentration.
The initial settings can also be completed simply
by following the Wizard procedures.
Use existing calibration curves for measurements. Useful for sample
concentration order checks and for checking the instrument sensitivity.
One-touch re-measurement operation. No need to repeat the
measurement from the beginning.
Start Wizard.
Create and save multiple calibration curves on a single worksheet.
Select the appropriate calibration curve for the sample concentration and
composition.
Select element.
Set calibration curve and sample parameters.
Combining a high-performance autosampler simplifies advanced settings
Set monochromator parameters.
Real-time drying and ashing progress display for furnace measurement
Connect to instrument.
A higher temperature cleaning operation than the cleaning stage can be
added for each measurement. This reduces the graphite tube memory
effect. Tube deterioration is reduced by minimizing the cleaning
operations.
Measurement Screen Layout Shows Measurement Status at a Glance
Measured element
Real-time signal monitor
Signal profile display
Calibration curve display
Add optional autosampler rinsing operations to ensure adequate rinsing
of the autosampler after completing the measurement of
high-concentration samples.
Set up sample dilution and spike-and-recovery tests on the same screen.
MRT (Measured Results Table)
The worksheet shows sample names,
absorbance, concentrations,
and correction calibration results.
Overlay function allows size and shape comparison of peak profiles.
Large-volume injection permits high-sensitivity measurements.
Precision control of drying and ashing temperatures reduces analysis time.
Lamp History Function Assists with Lamp Management
The accumulated operation time for each lamp displayed on the lamp
registration screen assists with lamp service life management. Multiple lamps
for the same element are differentiated using lamp IDs.
16
17
Supports System Management and Accuracy Management
Accuracy Management (QA/QC) Functions Permit User Level Management, Audit Trail, and Electronic Signatures
FDA 21 CFR Part 11 Compliance
A combination of enhanced WizAArd software with Shimadzu's network-compatible CLASS-Agent data
management software supports FDA 21 CFR Part 11 Compliance.
QA/QC compatibility
The QA/QC functions permit accuracy management.
WizAArd used alone provides comprehensive system policy setting, user management, log browser, audit trail,
and electronic signature system management functions.
Hardware validation software installed as standard.
Comprehensive Data Management Functions
System Policy
CLASS-Agent (Option)
The system policy settings allow advanced
password and lockout settings. The levels
can be set in stages from "unrestricted" to
"Part 11 compatible".
User Management
User management restricts the users of the
system. It registers each user to determine
how each one uses the software.
Connection to CLASS-Agent permits
efficient, long-term database management
of large amounts of analysis data.
Hardware validation software installed as standard
The hardware validation software makes it simple to evaluate the instrument performance.
When combined with the autosampler, it automatically determines the wavelength accuracy,
noise level, baseline drift, absorbance and repeatability, and prints out the results of
comparison with the pass criteria.
Log Browser
The log browser easily displays a variety of logs to check the
system modification history and other log information.
* Data from all Shimadzu analytical instruments can be centrally managed by CLASS-Agent.
18
19
Supports System Management and Accuracy Management
Accuracy Management (QA/QC) Functions Permit User Level Management, Audit Trail, and Electronic Signatures
FDA 21 CFR Part 11 Compliance
A combination of enhanced WizAArd software with Shimadzu's network-compatible CLASS-Agent data
management software supports FDA 21 CFR Part 11 Compliance.
QA/QC compatibility
The QA/QC functions permit accuracy management.
WizAArd used alone provides comprehensive system policy setting, user management, log browser, audit trail,
and electronic signature system management functions.
Hardware validation software installed as standard.
Comprehensive Data Management Functions
System Policy
CLASS-Agent (Option)
The system policy settings allow advanced
password and lockout settings. The levels
can be set in stages from "unrestricted" to
"Part 11 compatible".
User Management
User management restricts the users of the
system. It registers each user to determine
how each one uses the software.
Connection to CLASS-Agent permits
efficient, long-term database management
of large amounts of analysis data.
Hardware validation software installed as standard
The hardware validation software makes it simple to evaluate the instrument performance.
When combined with the autosampler, it automatically determines the wavelength accuracy,
noise level, baseline drift, absorbance and repeatability, and prints out the results of
comparison with the pass criteria.
Log Browser
The log browser easily displays a variety of logs to check the
system modification history and other log information.
* Data from all Shimadzu analytical instruments can be centrally managed by CLASS-Agent.
18
19
Standard Parts / PC / Software
Common to All AA-7000 Series Models
Part Name
Optional Accessories / Related Products
Quantity
Cat. No.
Cable set
1
071-60816-12
Cards (set of 10)
1
206-52046-91
Instruction manual
1
206-97176
Inspection report
1
206-77551-02
Safety inspection sheet (safety instruction)
1
206-97225
1
228-30164
1
037-70238-01
1
221-40500
3
206-57770
Declaration of conformity, explanatory notes (ISO-9001)
Warranty label
Registration form
Serial number label
* The PC, monitor, printer, hollow cathode lamps, high-temperature burner head, pressure regulator and compressor are not included in the standard configuration.
For Flame Analysis
Part Name
High-temperature burner head
Sample platform
206-77655-91
Air compressor
208-91753-91
100 VAC, 50/60 Hz, with mist separator
208-91750-31
100 VAC, 50/60 Hz, with mist separator
206-52458-91
Required if using an air compressor other than above.
040-72020-01
For C2H2
040-72019-11
For dinitrogen oxide gas
Hose ASSY (for air supply)
Micro sampling kit
206-77540-91
Required to use the flame micro sampling method. ASC-7000 and
ASK-7000 (or ASC stand kit) are also required.
Hose ASSY (for C2H2 gas supply)
Hose clamp (16 mm)
O-ring set
206-77620-92
O-ring set for organic solvents
Cleaning wire
Sampling tube (PTFE)
Sampling tube (for organic solvent samples)
MAF-85S compressed gas regulator
1
206-50389-91
1
206-50389-92
2
037-61019
1
201-79229-01
2
204-05899-01
1
206-50772-91
Polyethylene tubing, No. 3
1 ( 0.3m )
Polyethylene tubing, 8 Г— 1 (drain tubing)
1 ( 2.4m )
Drain ASSY
Grease (in cup)
Funnel
Declaration of conformity, English version (ISO-9001)
Cat. No.
200-31328-01
016-43201-02
1
206-77413-91
1
206-50442-91
1
206-77243-91
1
206-84934-66
2
206-77617-91
YR-71 compressed gas regulator
Quantity
.Extremely corrosion-resistant
.Can also be used for Air-C H flame.
Flow meter kit
Mist separator kit
Part Name
Remarks
Made of pure titanium. Air-cooled. 5 cm slot for N2O-C2H2 flame
2
Low-noise air compressor
AA-7000F Standard Parts List
Cat. No.
206-77530-91
For Furnace Analysis
Float-type flow meter for support gas
Dimensions: 250 W Г— 130 D Г— 170 H mm
Vial mounting platform: 220 W Г— 95 D mm
Vial positions: 5-step switching
Part Name
Cat. No.
GFA-7000 graphite furnace atomizer
206-77700-XX
Remarks
High-density graphite tube
206-50587
Pyro-coated graphite tube
206-50588
Platform tube
206-50887-02
MAF-106S compressed gas regulator
040-72019-21
Argon gas cylinder
CA-1112 cooling water circulation unit
044-01809-03
For cooling GFA, 100 VAC, 50/60 Hz
Cooler connection kit
206-84373-91
Cooling water tube ASSY
206-51028-91
Connecting tubes when using tap water to cool GFA
Regulator ASSY
206-86147-91
Decompression valve when using tap water to cool GFA
ISpecifications
Heating control system: Drying: Digital current control
(with automatic temperature calibration function)
Ashing, atomization: Digital temperature control via
optical sensor
Heating temperature range: Ambient to 3000 В°C
Inner gas type: Dual automatic switching type
Inner gas flow rate: 0 to 1.50 L/min, 0.01 L/min increments
Ar gas: 3.5 L/min max.
Cooling water: Cooling water circulation unit or tap for process water
Water temperature: 10 to 30 В°C, flow rate: 0.6 to 1.5 L/min
* Separately order the parts below.
AA-7000G Standard Parts List
Part Name
Declaration of conformity, English version (ISO-9001)
Furnace mounting plate
Quantity
Cat. No.
1
206-84934-67
1
206-77704
* GFA-7000 is not included in the standard configuration.
Personal Computer / Monitor
Operating system
Microsoft Windows Vista Business or XP Professional
CPU
Intel В® Celeron 420 (1.60В GHz) or higher
RAM
1 GB or higher (Vista) or 512В MB or higher (XP)
Monitor
XGA (1024В Г—В 768 dots) or higher
Storage device
One CD-ROM drive (for installing software) Requires 60В MB min. hard disk space for installation
I/O port
One serial port (for AA control)
Parts for cooling water connections
Autosampler
Part Name
ASC-7000 autosampler
Select one of the three types of graphite tubes according to the aim of the analysis.
Operational temperature range: -10 to +40 В°C
For connecting GFA and CA-1112
Cat. No.
206-77600-XX
Remarks
ISpecifications
System: Flame / furnace
Function: Zero-point detection, auto rinse, auto diagnosis, random access
Maximum reagent / sample positions: Reagents: 8 positions, samples: 60 positions
* Separately order the parts below.
* Windows is a registered trademark of Microsoft Corp., U.S.A.
For flame analysis
20
For furnace analysis
ASC stand kit
206-77650-91
Required to mount ASC-7000 to conduct flame analysis only.
ASK-7000 extension unit for furnace analysis
206-77550-91
Permits both flame and furnace analysis.
Nozzle ASSY, HVG
206-67563
Required to use ASC-7000 and HVG-1.
21
Standard Parts / PC / Software
Common to All AA-7000 Series Models
Part Name
Optional Accessories / Related Products
Quantity
Cat. No.
Cable set
1
071-60816-12
Cards (set of 10)
1
206-52046-91
Instruction manual
1
206-97176
Inspection report
1
206-77551-02
Safety inspection sheet (safety instruction)
1
206-97225
1
228-30164
1
037-70238-01
1
221-40500
3
206-57770
Declaration of conformity, explanatory notes (ISO-9001)
Warranty label
Registration form
Serial number label
* The PC, monitor, printer, hollow cathode lamps, high-temperature burner head, pressure regulator and compressor are not included in the standard configuration.
For Flame Analysis
Part Name
High-temperature burner head
Sample platform
206-77655-91
Air compressor
208-91753-91
100 VAC, 50/60 Hz, with mist separator
208-91750-31
100 VAC, 50/60 Hz, with mist separator
206-52458-91
Required if using an air compressor other than above.
040-72020-01
For C2H2
040-72019-11
For dinitrogen oxide gas
Hose ASSY (for air supply)
Micro sampling kit
206-77540-91
Required to use the flame micro sampling method. ASC-7000 and
ASK-7000 (or ASC stand kit) are also required.
Hose ASSY (for C2H2 gas supply)
Hose clamp (16 mm)
O-ring set
206-77620-92
O-ring set for organic solvents
Cleaning wire
Sampling tube (PTFE)
Sampling tube (for organic solvent samples)
MAF-85S compressed gas regulator
1
206-50389-91
1
206-50389-92
2
037-61019
1
201-79229-01
2
204-05899-01
1
206-50772-91
Polyethylene tubing, No. 3
1 ( 0.3m )
Polyethylene tubing, 8 Г— 1 (drain tubing)
1 ( 2.4m )
Drain ASSY
Grease (in cup)
Funnel
Declaration of conformity, English version (ISO-9001)
Cat. No.
200-31328-01
016-43201-02
1
206-77413-91
1
206-50442-91
1
206-77243-91
1
206-84934-66
2
206-77617-91
YR-71 compressed gas regulator
Quantity
.Extremely corrosion-resistant
.Can also be used for Air-C H flame.
Flow meter kit
Mist separator kit
Part Name
Remarks
Made of pure titanium. Air-cooled. 5 cm slot for N2O-C2H2 flame
2
Low-noise air compressor
AA-7000F Standard Parts List
Cat. No.
206-77530-91
For Furnace Analysis
Float-type flow meter for support gas
Dimensions: 250 W Г— 130 D Г— 170 H mm
Vial mounting platform: 220 W Г— 95 D mm
Vial positions: 5-step switching
Part Name
Cat. No.
GFA-7000 graphite furnace atomizer
206-77700-XX
Remarks
High-density graphite tube
206-50587
Pyro-coated graphite tube
206-50588
Platform tube
206-50887-02
MAF-106S compressed gas regulator
040-72019-21
Argon gas cylinder
CA-1112 cooling water circulation unit
044-01809-03
For cooling GFA, 100 VAC, 50/60 Hz
Cooler connection kit
206-84373-91
Cooling water tube ASSY
206-51028-91
Connecting tubes when using tap water to cool GFA
Regulator ASSY
206-86147-91
Decompression valve when using tap water to cool GFA
ISpecifications
Heating control system: Drying: Digital current control
(with automatic temperature calibration function)
Ashing, atomization: Digital temperature control via
optical sensor
Heating temperature range: Ambient to 3000 В°C
Inner gas type: Dual automatic switching type
Inner gas flow rate: 0 to 1.50 L/min, 0.01 L/min increments
Ar gas: 3.5 L/min max.
Cooling water: Cooling water circulation unit or tap for process water
Water temperature: 10 to 30 В°C, flow rate: 0.6 to 1.5 L/min
* Separately order the parts below.
AA-7000G Standard Parts List
Part Name
Declaration of conformity, English version (ISO-9001)
Furnace mounting plate
Quantity
Cat. No.
1
206-84934-67
1
206-77704
* GFA-7000 is not included in the standard configuration.
Personal Computer / Monitor
Operating system
Microsoft Windows Vista Business or XP Professional
CPU
Intel В® Celeron 420 (1.60В GHz) or higher
RAM
1 GB or higher (Vista) or 512В MB or higher (XP)
Monitor
XGA (1024В Г—В 768 dots) or higher
Storage device
One CD-ROM drive (for installing software) Requires 60В MB min. hard disk space for installation
I/O port
One serial port (for AA control)
Parts for cooling water connections
Autosampler
Part Name
ASC-7000 autosampler
Select one of the three types of graphite tubes according to the aim of the analysis.
Operational temperature range: -10 to +40 В°C
For connecting GFA and CA-1112
Cat. No.
206-77600-XX
Remarks
ISpecifications
System: Flame / furnace
Function: Zero-point detection, auto rinse, auto diagnosis, random access
Maximum reagent / sample positions: Reagents: 8 positions, samples: 60 positions
* Separately order the parts below.
* Windows is a registered trademark of Microsoft Corp., U.S.A.
For flame analysis
20
For furnace analysis
ASC stand kit
206-77650-91
Required to mount ASC-7000 to conduct flame analysis only.
ASK-7000 extension unit for furnace analysis
206-77550-91
Permits both flame and furnace analysis.
Nozzle ASSY, HVG
206-67563
Required to use ASC-7000 and HVG-1.
21
Optional Accessories / Related Products
Dual Atomizer System
Part Name
AAC-7000 auto atomizer changer
For High-Sensitivity Hg Analysis
Cat. No.
Remarks
206-77701-91
Required to add GFA-7000 to AA-7000F to conduct furnace analysis.
Permits automatic burner position setting for flame analysis with AA-7000F.
Part Name
Cat. No.
MVU-1A mercury vaporizer unit
206-21932-XX
Remarks
This mercury vaporizer unit permits analysis using the reduction vaporization ––
atomic absorption method. It permits easy, high-sensitivity water quality
analysis.
ISpecifications
Hollow Cathode Lamps
Part Name
Cat. No.
Vaporization method: reduction vaporization with a reducing agent
Measurement method: recirculation
Flow cell: optical path length 100 mm (with quartz window)
Sample volume: 250 mL max.
Exhaust contamination prevention: adsorption trapping in mercury trap bottle
Dimensions: 200 W Г— 288 D Г— 287 H mm
Weight: Approx. 10 kg
Remarks
L-233 series
200-38422-XX
L-2433 series
200-38456-XX
For SR method
IStandard Major Items
Other Accessories
Part Name
Cat. No.
Analog output cable
MVU-1A unit: 1
Reaction vial (P/N: 200-93018): 5
Reaction vial lid (P/N: 204-21989): 2
Stirrer tip (P/N: 046-00617-06): 10
Mercury trap bottle (P/N: 038-00218): 1
Remarks
206-77707-91
Used for analog output to a pen recorder, etc.
One cable required per channel (atomic absorption/energy signal, background signal).
*Separately order the parts below.
. Gas flow cell (P/N: 201-98687)
. Holder for gas flow cell (P/N: 206-77703-91)
. Hg hollow cathode lamp (P/N: 200-38422-28)
For High-Sensitivity As, Se, Sb Analysis
Part Name
Cat. No.
HVG-1 hydride vapor generator
206-17143-XX
Remarks
Environmental standards prescribe the hydride generation method as one
method of As, Se, and Sb analysis.
1. Used with the AA-7000 Series, HVG-1 permits the rapid and accurate
quantitation of elements such as As, Se, Sb, Sn, Te, and Bi at several-ppb
levels.
2. Used with an ASC-7000 autosampler, it permits the automated serial analysis
of up to 60 samples.
* Nozzle ASSY, HVG (P/N: 206-67563) is required to use HVG-1 with an ASC-7000 Series
instrument. Order separately.
ISpecifications
Measurement method: continuous flow
Sample consumption: 0 to 7 mL/min, variable
Reagent consumption: 0 to 2.5 mL/min, variable
Atomizer: Heated absorption cell (heated by Air-C2H2 flame in standard system)
Carrier gas: Ar pressure: 3.2 kg/cm2, consumption: 70 mL/min
Power requirements: 100, 115, 220, 240 VAC, 30 VA, 50/60 Hz
Dimensions: 340 W Г— 220 D Г— 200 H mm
ICP Emission Spectrometers
ICPE-9000
Remarks
Offers both high throughput and easy analysis. Does not require processing of
large data volumes, such as spectral line selection, which was problematic with
the conventional Multi-ICP. Various assistant functions allow anyone to rapidly
perform accurate analysis.
Light source: Axial observation, compatible with mini torch
Monochromator / detector: Echelle CCD
Wavelength range: 167 to 800 nm (using vacuum monochromator)
IStandard Major Items
Hydride vapor generator, absorption cell (P/N: 200-66584-03)
Reagent bottles (P/N: 204-09363-01), gas hose, drain tube, etc.
For Higher Sensitivity with the Hydride Generation Method
Part Name
Cat. No.
SARF-16C atomic muffle furnace
(Electronic Cell Heater)
Remarks
208-97249
1. This dedicated furnace permits higher sensitivity measurements using the
hydride vapor generator than the flame heating method.
2. The temperature controller provides optimal control of the quartz cell
temperature.
3. Good temperature reproducibility. Extends cell life by preventing damage due
to overheating.
* The atomic muffler cannot be used when an AAC-7000 is installed on an AA-7000F.
Temperature sensor
Cell heater
HVG-1 hydride vapor generator
Controller
Atomic absorption
spectrophotometer
ISpecifications
Furnace
Dimensions: 170 W Г— 110 D Г— 110 H mm
I.D.: Гё25 Г— 170 mm
Heater: PYROMAX DS
Controller
Operational temperature range: Ambient to 1100 В°C
Temperature regulation range: В±2.5 В°C
Load capacity: 1000 W max.
Power requirements: 100 VAC, 400 VA, 50/60 Hz
Computer
* Use a stepdown transformer in regions that do not have a 100 V power supply.
Mounting adaptor
22
206-52135-91
For AA-7000F
206-83755-91
For AA-7000G
23
Optional Accessories / Related Products
Dual Atomizer System
Part Name
AAC-7000 auto atomizer changer
For High-Sensitivity Hg Analysis
Cat. No.
Remarks
206-77701-91
Required to add GFA-7000 to AA-7000F to conduct furnace analysis.
Permits automatic burner position setting for flame analysis with AA-7000F.
Part Name
Cat. No.
MVU-1A mercury vaporizer unit
206-21932-XX
Remarks
This mercury vaporizer unit permits analysis using the reduction vaporization ––
atomic absorption method. It permits easy, high-sensitivity water quality
analysis.
ISpecifications
Hollow Cathode Lamps
Part Name
Cat. No.
Vaporization method: reduction vaporization with a reducing agent
Measurement method: recirculation
Flow cell: optical path length 100 mm (with quartz window)
Sample volume: 250 mL max.
Exhaust contamination prevention: adsorption trapping in mercury trap bottle
Dimensions: 200 W Г— 288 D Г— 287 H mm
Weight: Approx. 10 kg
Remarks
L-233 series
200-38422-XX
L-2433 series
200-38456-XX
For SR method
IStandard Major Items
Other Accessories
Part Name
Cat. No.
Analog output cable
MVU-1A unit: 1
Reaction vial (P/N: 200-93018): 5
Reaction vial lid (P/N: 204-21989): 2
Stirrer tip (P/N: 046-00617-06): 10
Mercury trap bottle (P/N: 038-00218): 1
Remarks
206-77707-91
Used for analog output to a pen recorder, etc.
One cable required per channel (atomic absorption/energy signal, background signal).
*Separately order the parts below.
. Gas flow cell (P/N: 201-98687)
. Holder for gas flow cell (P/N: 206-77703-91)
. Hg hollow cathode lamp (P/N: 200-38422-28)
For High-Sensitivity As, Se, Sb Analysis
Part Name
Cat. No.
HVG-1 hydride vapor generator
206-17143-XX
Remarks
Environmental standards prescribe the hydride generation method as one
method of As, Se, and Sb analysis.
1. Used with the AA-7000 Series, HVG-1 permits the rapid and accurate
quantitation of elements such as As, Se, Sb, Sn, Te, and Bi at several-ppb
levels.
2. Used with an ASC-7000 autosampler, it permits the automated serial analysis
of up to 60 samples.
* Nozzle ASSY, HVG (P/N: 206-67563) is required to use HVG-1 with an ASC-7000 Series
instrument. Order separately.
ISpecifications
Measurement method: continuous flow
Sample consumption: 0 to 7 mL/min, variable
Reagent consumption: 0 to 2.5 mL/min, variable
Atomizer: Heated absorption cell (heated by Air-C2H2 flame in standard system)
Carrier gas: Ar pressure: 3.2 kg/cm2, consumption: 70 mL/min
Power requirements: 100, 115, 220, 240 VAC, 30 VA, 50/60 Hz
Dimensions: 340 W Г— 220 D Г— 200 H mm
ICP Emission Spectrometers
ICPE-9000
Remarks
Offers both high throughput and easy analysis. Does not require processing of
large data volumes, such as spectral line selection, which was problematic with
the conventional Multi-ICP. Various assistant functions allow anyone to rapidly
perform accurate analysis.
Light source: Axial observation, compatible with mini torch
Monochromator / detector: Echelle CCD
Wavelength range: 167 to 800 nm (using vacuum monochromator)
IStandard Major Items
Hydride vapor generator, absorption cell (P/N: 200-66584-03)
Reagent bottles (P/N: 204-09363-01), gas hose, drain tube, etc.
For Higher Sensitivity with the Hydride Generation Method
Part Name
Cat. No.
SARF-16C atomic muffle furnace
(Electronic Cell Heater)
Remarks
208-97249
1. This dedicated furnace permits higher sensitivity measurements using the
hydride vapor generator than the flame heating method.
2. The temperature controller provides optimal control of the quartz cell
temperature.
3. Good temperature reproducibility. Extends cell life by preventing damage due
to overheating.
* The atomic muffler cannot be used when an AAC-7000 is installed on an AA-7000F.
Temperature sensor
Cell heater
HVG-1 hydride vapor generator
Controller
Atomic absorption
spectrophotometer
ISpecifications
Furnace
Dimensions: 170 W Г— 110 D Г— 110 H mm
I.D.: Гё25 Г— 170 mm
Heater: PYROMAX DS
Controller
Operational temperature range: Ambient to 1100 В°C
Temperature regulation range: В±2.5 В°C
Load capacity: 1000 W max.
Power requirements: 100 VAC, 400 VA, 50/60 Hz
Computer
* Use a stepdown transformer in regions that do not have a 100 V power supply.
Mounting adaptor
22
206-52135-91
For AA-7000F
206-83755-91
For AA-7000G
23
Specifications
Autosampler
Main Unit
Basics
Optics
Control
RS-232C communication control from AA unit
Maximum reagent / sample positions
Reagents: 8 positions
0.2, 0.7, 1.3, 2.0L nm (4-step automatic switching)
Rinse water bottle
2L
Detector
Photomultiplier tube
Optics
Power requirements
Choose from 100, 120, 220, or 230 VAC, 50 VA, 50/60 Hz
Optical double-beam
185.0 to 900.0 nm
Monochromator
Aberration-corrected Czerny-Turner mounting
Bandwidth
Background correction method
Lamp mode
EMISSION, NON-BGC, BGC-D2, BGC-SR
Software requirements
Microsoft Windows Vista Business / XP Professional
Parameter setting
Wizard method
Measurement mode
Flame continuous method, flame micro sampling method, furnace method, flame emission method
. Calibration curve method (select primary, secondary, tertiary) . Standard addition method, simple standard addition method (primary expression)
Repeat analysis
Up to 20 repetitions. Mean value, standard deviation (SD) and coefficient of variation (RSD) display
Automatic exclusion of deviant values by setting SD and %RSD
Baseline correction
Automatic correction of baseline drift by offset correction in peak height / peak area modes.
Signal processing segment setting
Signal processing segments can be changed in peak height / peak area modes.
Flame Continuous Method
Samples: 60 positions
(random access available with reagents or samples)
Functions
Zero-point detection, auto diagnosis, auto rinse, random access
Sample volume
Sample vials: 15 mL, reagent vials: 53 mL
Nozzle rinse
Solvent rinse aspiration method
Dimensions and weight
Sampler: 340 W Г— 280 D Г— 270 H mm, 8 kg
Controller (including stand): 260 W Г— 200 D Г— 320 H mm, 7 kg
Furnace Method / Flame Micro Sampling Method
Functions
Zero-point detection, auto diagnosis, auto rinse, auto mixing, random access
Sensitivity correction
Automatic calibration curve correction function using sensitivity monitoring
Sample volume
Sample vials: 2 mL, reagent vials: 20 mL
Nozzle rinse
Sampling functions
Solvent rinse discharge method
Analog output
2 channels (atomic absorption/energy signal, background signal)
Output range: 5.0, 2.5, 1.25, 0.625 Abs./V (each settable in 4 stages)
Fixed at 1 V F.S. in EMISSION mode.
Tabular data processing
Final concentration calculations based on sampled volume, dilution rate, fixed volume, and factor inputs
Syringe
Injection volume
Dilution function, reagent addition function
250 ОјL
2 to 90 ОјL
Template functions available
Repeatability
Procedure/result display
MRT (Measurement Results Table) worksheet
Report generation
Summary report
Carryover
Rinse port: Less than 0.0001
Mixing port: Less than 0.00001
QA/QC
Select whether to continue or discontinue measurements based on results of evaluation on coefficient of correlation,
%RSD, ICV . ICB, CCV . CCB, PB, LCS, SPK, PDS, and DUP.
Mixing port rinse
Solvent rinse discharge method, solvent discharge and rinse with next sample
Re-analysis
Digital recording
Power requirements
Dimensions and weight
Ambient temperature / humidity
Flame
. Select whether on not to conduct re-analysis.
. Automatic dilution and re-analysis of unknown samples via autosampler (flame micro sampling method, furnace method)
. Management by login ID and password . Control user access authority by user level
. Log record . Audit trail . Electronic signatures
..
Choose from 100, 120, 220, or 230 VAC, 50/60 Hz (Power is required separately for the personal computer.)
AA-7000F: 700 W Г— 588 D Г— 714 H mm, 72 kg
AA-7000G: 700 W Г— 580 D Г— 538 H mm, 65 kg
(Protruding parts and optional equipment are not included.)
10 to 35 В°C, 20 to 80% (less than 70% when temperature is higher than 30 В°C)
.
.
.
.
.
Type
Air-cooled pre-mix type
Burner head
Titanium 10 cm slot (5 cm titanium slot for N2O-C2H2 flame available as an option)
Nebulizer
Chamber
Pt-lr capillary
PTFE orifice
Engineering plastics
ceramic impact bead (capable of handling hydrofluoric acid)
. Lateral/vertical manual adjustment
AA-7000F/AA-7000G
Positioning
Angle adjustment
Flow rate control
Gas
control
BGC-SR (high-speed self-reversal method) (185.0 to 900.0 nm)
BGC-D2 (D2 lamp method) (185.0 to 430.0 nm)
6-lamp turret, 2 lamps simultaneously lit (1 for measurement, 1 warming up for next measurement)
Recall of parameters
Burner
unit
..
Number of HC lamps
Concentration computation mode
Data
processing
Common Specifications
Wavelength range
Safety measures
Furnace (GFA-7000)
Heating temperature range
Heating control system
Setting heating conditions
Safety measures
. Automatic flame/furnace switching by motor . Automatic search of optimum burner height
AA-7000F (with an optional auto atomizer changer mounted)
0 to 90В° (Angle adjustment is not possible if the optional auto atomizer changer and GFA-7000 are both installed.)
. Automatic fuel gas flow rate setting (0.1 L/min step) . Automatic search of optimum gas flow rate
. Automatic gas leak check . Automatic Air-N O switching as C H flow rate increases . Flame monitor . Prevention of wrong burner head use
. Gas pressure monitor . Drain tank level monitor . Automatic flame extinction upon power outage or sudden power interruption
. Automatic flame extinction via flame vibration sensor . Internal fan stop sensor
2
2
2
. Ambient to 3,000 В°C
. Drying: Digital current control with automatic temperature calibration function
. Ashing, Atomization: Digital temperature control via optical sensor
. Maximum 20 stages . Heating mode: RAMP/STEP . Inner gas type: Dual automatic switching type
. High-sensitivity mode setting . Enrichment in furnace: Maximum 20 times
. Optimum temperature program search support function . Inner gas flow rate: 0 to 1.50 L/min
. Cooling water flow rate monitor . Gas pressure monitor . Overcurrent protection unit (double check by circuit protector and optical sensor)
. Furnace block cooling check
.
.
Mixing function
Maximum number of
mixing reagents added
Auto dilution / re-analysis
Dimensions and weight
1%RSD (20 ОјL)
Performed in mixing port. Maximum mixture volume: 0.6 mL
Up to 4 solutions
Number of solutions possible for mixing
. Calibration curve method: 5 solutions max. (sample + 4 types of reagents)
. Standard addition method: 6 solutions max. (sample + standard solution + 4 types of reagents)
For measurement result on unknown samples:
. If extrapolation of calibration curve is possible: automatic calculation of dilution rate and dilution to bring concentration within calibration curve range
. If extrapolation of calibration curve is not possible: dilution rate fixed at 10x
Sampler (including sliding parts): 340 W Г— 280 D Г— 400 H mm, 11 kg
Controller: 235 W Г— 110 D Г— 240 H mm (built into GFA-7000), 6 kg
Required Software Operation Environment
Shimadzu recommends the PC and monitor in the PC Set for AA-7000 Series with WizAArd Software Pre-Installed.
Purchase a printer separately. The personal computer and monitor specifications are as follows.
Operating system
CPU
RAM
Monitor
Storage device
I/O port
Peripheral device
Microsoft Windows Vista Business / XP Professional
IntelВ® Celeron 420 (1.60 GHz or higher)
1 GB or higher (Vista) or 512 MB or higher (XP)
XGA (1024 Г— 768 dots) or higher
One CD-ROM drive (for installing a software)
Requires 60 MB min. hard disk space for AA installation.
One serial port (for AA control)
Monitor, keyboard, mouse, printer
* Notes
. The PC, monitor and printer are not included in the standard configuration. Prepare them separately.
. The ASC-7000, ASK-7000 and GFA-7000 are optional accessories.
. The separately available micro sampling kit is required for the flame micro sampling method.
. Microsoft Windows Vista Business or XP Professional is required to run the AA-7000 Series software.
. Microsoft Windows Vista Business and XP Professional are registered trademarks of Microsoft Corp., U.S.A.
. The AA-7000 Series does not conform to IEC60601 –– Medical Electrical Equipment.
. Lateral/vertical manual adjustment
AA-7000F (with an optional auto atomizer changer mounted)
. Automatic flame/furnace switching by motor
AA-7000F/AA-7000G
Positioning
24
Power requirements
200, 220, 230, or 240 VAC В±5%, 6000 VA, 50/60 Hz
Dimensions and weight
260 W Г— 560 D Г— 510 H mm, 46 kg
25
Specifications
Autosampler
Main Unit
Basics
Optics
Control
RS-232C communication control from AA unit
Maximum reagent / sample positions
Reagents: 8 positions
0.2, 0.7, 1.3, 2.0L nm (4-step automatic switching)
Rinse water bottle
2L
Detector
Photomultiplier tube
Optics
Power requirements
Choose from 100, 120, 220, or 230 VAC, 50 VA, 50/60 Hz
Optical double-beam
185.0 to 900.0 nm
Monochromator
Aberration-corrected Czerny-Turner mounting
Bandwidth
Background correction method
Lamp mode
EMISSION, NON-BGC, BGC-D2, BGC-SR
Software requirements
Microsoft Windows Vista Business / XP Professional
Parameter setting
Wizard method
Measurement mode
Flame continuous method, flame micro sampling method, furnace method, flame emission method
. Calibration curve method (select primary, secondary, tertiary) . Standard addition method, simple standard addition method (primary expression)
Repeat analysis
Up to 20 repetitions. Mean value, standard deviation (SD) and coefficient of variation (RSD) display
Automatic exclusion of deviant values by setting SD and %RSD
Baseline correction
Automatic correction of baseline drift by offset correction in peak height / peak area modes.
Signal processing segment setting
Signal processing segments can be changed in peak height / peak area modes.
Flame Continuous Method
Samples: 60 positions
(random access available with reagents or samples)
Functions
Zero-point detection, auto diagnosis, auto rinse, random access
Sample volume
Sample vials: 15 mL, reagent vials: 53 mL
Nozzle rinse
Solvent rinse aspiration method
Dimensions and weight
Sampler: 340 W Г— 280 D Г— 270 H mm, 8 kg
Controller (including stand): 260 W Г— 200 D Г— 320 H mm, 7 kg
Furnace Method / Flame Micro Sampling Method
Functions
Zero-point detection, auto diagnosis, auto rinse, auto mixing, random access
Sensitivity correction
Automatic calibration curve correction function using sensitivity monitoring
Sample volume
Sample vials: 2 mL, reagent vials: 20 mL
Nozzle rinse
Sampling functions
Solvent rinse discharge method
Analog output
2 channels (atomic absorption/energy signal, background signal)
Output range: 5.0, 2.5, 1.25, 0.625 Abs./V (each settable in 4 stages)
Fixed at 1 V F.S. in EMISSION mode.
Tabular data processing
Final concentration calculations based on sampled volume, dilution rate, fixed volume, and factor inputs
Syringe
Injection volume
Dilution function, reagent addition function
250 ОјL
2 to 90 ОјL
Template functions available
Repeatability
Procedure/result display
MRT (Measurement Results Table) worksheet
Report generation
Summary report
Carryover
Rinse port: Less than 0.0001
Mixing port: Less than 0.00001
QA/QC
Select whether to continue or discontinue measurements based on results of evaluation on coefficient of correlation,
%RSD, ICV . ICB, CCV . CCB, PB, LCS, SPK, PDS, and DUP.
Mixing port rinse
Solvent rinse discharge method, solvent discharge and rinse with next sample
Re-analysis
Digital recording
Power requirements
Dimensions and weight
Ambient temperature / humidity
Flame
. Select whether on not to conduct re-analysis.
. Automatic dilution and re-analysis of unknown samples via autosampler (flame micro sampling method, furnace method)
. Management by login ID and password . Control user access authority by user level
. Log record . Audit trail . Electronic signatures
..
Choose from 100, 120, 220, or 230 VAC, 50/60 Hz (Power is required separately for the personal computer.)
AA-7000F: 700 W Г— 588 D Г— 714 H mm, 72 kg
AA-7000G: 700 W Г— 580 D Г— 538 H mm, 65 kg
(Protruding parts and optional equipment are not included.)
10 to 35 В°C, 20 to 80% (less than 70% when temperature is higher than 30 В°C)
.
.
.
.
.
Type
Air-cooled pre-mix type
Burner head
Titanium 10 cm slot (5 cm titanium slot for N2O-C2H2 flame available as an option)
Nebulizer
Chamber
Pt-lr capillary
PTFE orifice
Engineering plastics
ceramic impact bead (capable of handling hydrofluoric acid)
. Lateral/vertical manual adjustment
AA-7000F/AA-7000G
Positioning
Angle adjustment
Flow rate control
Gas
control
BGC-SR (high-speed self-reversal method) (185.0 to 900.0 nm)
BGC-D2 (D2 lamp method) (185.0 to 430.0 nm)
6-lamp turret, 2 lamps simultaneously lit (1 for measurement, 1 warming up for next measurement)
Recall of parameters
Burner
unit
..
Number of HC lamps
Concentration computation mode
Data
processing
Common Specifications
Wavelength range
Safety measures
Furnace (GFA-7000)
Heating temperature range
Heating control system
Setting heating conditions
Safety measures
. Automatic flame/furnace switching by motor . Automatic search of optimum burner height
AA-7000F (with an optional auto atomizer changer mounted)
0 to 90В° (Angle adjustment is not possible if the optional auto atomizer changer and GFA-7000 are both installed.)
. Automatic fuel gas flow rate setting (0.1 L/min step) . Automatic search of optimum gas flow rate
. Automatic gas leak check . Automatic Air-N O switching as C H flow rate increases . Flame monitor . Prevention of wrong burner head use
. Gas pressure monitor . Drain tank level monitor . Automatic flame extinction upon power outage or sudden power interruption
. Automatic flame extinction via flame vibration sensor . Internal fan stop sensor
2
2
2
. Ambient to 3,000 В°C
. Drying: Digital current control with automatic temperature calibration function
. Ashing, Atomization: Digital temperature control via optical sensor
. Maximum 20 stages . Heating mode: RAMP/STEP . Inner gas type: Dual automatic switching type
. High-sensitivity mode setting . Enrichment in furnace: Maximum 20 times
. Optimum temperature program search support function . Inner gas flow rate: 0 to 1.50 L/min
. Cooling water flow rate monitor . Gas pressure monitor . Overcurrent protection unit (double check by circuit protector and optical sensor)
. Furnace block cooling check
.
.
Mixing function
Maximum number of
mixing reagents added
Auto dilution / re-analysis
Dimensions and weight
1%RSD (20 ОјL)
Performed in mixing port. Maximum mixture volume: 0.6 mL
Up to 4 solutions
Number of solutions possible for mixing
. Calibration curve method: 5 solutions max. (sample + 4 types of reagents)
. Standard addition method: 6 solutions max. (sample + standard solution + 4 types of reagents)
For measurement result on unknown samples:
. If extrapolation of calibration curve is possible: automatic calculation of dilution rate and dilution to bring concentration within calibration curve range
. If extrapolation of calibration curve is not possible: dilution rate fixed at 10x
Sampler (including sliding parts): 340 W Г— 280 D Г— 400 H mm, 11 kg
Controller: 235 W Г— 110 D Г— 240 H mm (built into GFA-7000), 6 kg
Required Software Operation Environment
Shimadzu recommends the PC and monitor in the PC Set for AA-7000 Series with WizAArd Software Pre-Installed.
Purchase a printer separately. The personal computer and monitor specifications are as follows.
Operating system
CPU
RAM
Monitor
Storage device
I/O port
Peripheral device
Microsoft Windows Vista Business / XP Professional
IntelВ® Celeron 420 (1.60 GHz or higher)
1 GB or higher (Vista) or 512 MB or higher (XP)
XGA (1024 Г— 768 dots) or higher
One CD-ROM drive (for installing a software)
Requires 60 MB min. hard disk space for AA installation.
One serial port (for AA control)
Monitor, keyboard, mouse, printer
* Notes
. The PC, monitor and printer are not included in the standard configuration. Prepare them separately.
. The ASC-7000, ASK-7000 and GFA-7000 are optional accessories.
. The separately available micro sampling kit is required for the flame micro sampling method.
. Microsoft Windows Vista Business or XP Professional is required to run the AA-7000 Series software.
. Microsoft Windows Vista Business and XP Professional are registered trademarks of Microsoft Corp., U.S.A.
. The AA-7000 Series does not conform to IEC60601 –– Medical Electrical Equipment.
. Lateral/vertical manual adjustment
AA-7000F (with an optional auto atomizer changer mounted)
. Automatic flame/furnace switching by motor
AA-7000F/AA-7000G
Positioning
24
Power requirements
200, 220, 230, or 240 VAC В±5%, 6000 VA, 50/60 Hz
Dimensions and weight
260 W Г— 560 D Г— 510 H mm, 46 kg
25
Installation Conditions
Dimensions
Make sure all required equipment is available before starting instrument installation. See the installation guidelines for details.
Wall
200
Table
150
Example of recommended piping for the atomic absorption system
588
Stop cock
Setting pressure 0.11 MPa
Stop cock
Gas tube
Humidity range
Withstand load: 200 kg *1, 2
Material
Dimensions
Gas supply
Air
pressure
(set secondary
Dinitrogen oxide
pressure)
Acetylene gas
Pressure range 0.3 MPa
Setting pressure 0.09 MPa
200 max.
Setting pressure 0.4 MPa
Dinitrogen
oxide gas
Pressure range 1 MPa
Setting pressure 0.35 MPa
Material
Exhaust duct
Dimensions
Intake capacity
Argon gas
Pressure range 1 MPa
Setting pressure 0.35 MPa
Гё150 to Гё200
Mist separator
Pressure range 1 MPa
Setting pressure 0.35 MPa
Lab
Exhaust fan
Outdoor
80 mm max.
Damper
Setting pressure 0.4 MPa
Air compressor
Stainless *3
7 mm min. I.D. Г— 80 mm min. (L)
Cylinder: 0.11 MPa
Lab: 0.09 MPa
Compressor: 0.4 MPa
Lab: 0.35 MPa
Cylinder: 0.4 MPa
Lab: 0.35 MPa
Cylinder: 0.4 MPa
Argon
Setting pressure 0.4 MPa
20 to 80% (less than 70% when temperature is higher than 30 В°C)
1500 mm min. (W) Г— 700 mm min. (D)
Acetylene
Pressure regulator
280
Temperature range 10 to 35 В°C
Testing bench
870
465
Stainless steel piping
(inner diameter greater than 7 mm)
Operating
environment
Cooling water (GFA-7000)
Lab: 0.35 MPa
Stainless
For flame: Approx. 500 mm W Г— 500 mm D
For furnace: Гё150 mm to Гё200 mm
For flame: 600 to 1200 m3/hr
For furnace: 10 to 180 m3/hr
Cooling water circulation unit or tap-water equipment *4
*1 Atomic absorption spectrophotometer stand recommended
*2 Maintain a free maintenance space of 150 to 200 mm to the sides and rear of the instrument.
*3 Do not use pipes containing copper, silver, gold, mercury (or alloys containing these metals) as
pipes for acetylene.
*4 If tap-water equipment is used, ensure that it meets the specifications below.
Compatible faucet
Faucet 13 or 12 to 15 mm O.D. rimmed faucet
Water temperature
10 to 30 В°C
Water flow rate
0.6 to 1.5 L/min
Supply pressure
0.08 to 0.15 MPa *5
Supply port
7 mm min.
*5 If the supply pressure exceeds 0.17 MPa, use the optional Regulator ASSY.
Гё150 to Гё200
Damper
Approx. 500 Г— 500
415
390
714
900 to 1000
1100 to 1200
415
340
700
960
Both for flame analysis and furnace analysis
960
For furnace analysis
Above dimensions do not include PC or printer.
26
Unit: mm
S=1: 2 0
27
Installation Conditions
Dimensions
Make sure all required equipment is available before starting instrument installation. See the installation guidelines for details.
Wall
200
Table
150
Example of recommended piping for the atomic absorption system
588
Stop cock
Setting pressure 0.11 MPa
Stop cock
Gas tube
Humidity range
Withstand load: 200 kg *1, 2
Material
Dimensions
Gas supply
Air
pressure
(set secondary
Dinitrogen oxide
pressure)
Acetylene gas
Pressure range 0.3 MPa
Setting pressure 0.09 MPa
200 max.
Setting pressure 0.4 MPa
Dinitrogen
oxide gas
Pressure range 1 MPa
Setting pressure 0.35 MPa
Material
Exhaust duct
Dimensions
Intake capacity
Argon gas
Pressure range 1 MPa
Setting pressure 0.35 MPa
Гё150 to Гё200
Mist separator
Pressure range 1 MPa
Setting pressure 0.35 MPa
Lab
Exhaust fan
Outdoor
80 mm max.
Damper
Setting pressure 0.4 MPa
Air compressor
Stainless *3
7 mm min. I.D. Г— 80 mm min. (L)
Cylinder: 0.11 MPa
Lab: 0.09 MPa
Compressor: 0.4 MPa
Lab: 0.35 MPa
Cylinder: 0.4 MPa
Lab: 0.35 MPa
Cylinder: 0.4 MPa
Argon
Setting pressure 0.4 MPa
20 to 80% (less than 70% when temperature is higher than 30 В°C)
1500 mm min. (W) Г— 700 mm min. (D)
Acetylene
Pressure regulator
280
Temperature range 10 to 35 В°C
Testing bench
870
465
Stainless steel piping
(inner diameter greater than 7 mm)
Operating
environment
Cooling water (GFA-7000)
Lab: 0.35 MPa
Stainless
For flame: Approx. 500 mm W Г— 500 mm D
For furnace: Гё150 mm to Гё200 mm
For flame: 600 to 1200 m3/hr
For furnace: 10 to 180 m3/hr
Cooling water circulation unit or tap-water equipment *4
*1 Atomic absorption spectrophotometer stand recommended
*2 Maintain a free maintenance space of 150 to 200 mm to the sides and rear of the instrument.
*3 Do not use pipes containing copper, silver, gold, mercury (or alloys containing these metals) as
pipes for acetylene.
*4 If tap-water equipment is used, ensure that it meets the specifications below.
Compatible faucet
Faucet 13 or 12 to 15 mm O.D. rimmed faucet
Water temperature
10 to 30 В°C
Water flow rate
0.6 to 1.5 L/min
Supply pressure
0.08 to 0.15 MPa *5
Supply port
7 mm min.
*5 If the supply pressure exceeds 0.17 MPa, use the optional Regulator ASSY.
Гё150 to Гё200
Damper
Approx. 500 Г— 500
415
390
714
900 to 1000
1100 to 1200
415
340
700
960
Both for flame analysis and furnace analysis
960
For furnace analysis
Above dimensions do not include PC or printer.
26
Unit: mm
S=1: 2 0
27
C122-E058A
AA-7000 Series
Shimadzu
Atomic Absorption Spectrophotometers
JQA-0376
Founded in 1875, Shimadzu Corporation, a leader in the
development of advanced technologies, has a distinguished
history of innovation built on the foundation of contributing to
society through science and technology. We maintain a global
network of sales, service, technical support and applications
centers on six continents, and have established long-term
relationships with a host of highly trained distributors located
in over 100 countries. For information about Shimadzu, and
to contact your local office, please visit our Web site at
www.shimadzu.com
SHIMADZU CORPORATION. International Marketing Division
3. Kanda-Nishikicho 1-chome, Chiyoda-ku, Tokyo 101-8448, Japan
Phone: 81(3)3219-5641 Fax. 81(3)3219-5710
URL http://www.shimadzu.com
Printed in Japan 3295-01910-30ANS
The contents of this brochure are subject to change without notice.